Specimen type | Serum, plasma, urine |
Sample volume | 20 ± 1μL |
Test time | 90 Seconds |
Sample capacity | single sample |
Unit | mOsm/kg H2O |
Resolution | 1 mOsm/kg H2O |
Sample throughput | can process 20 samples in less than 1 hour |
Range | 0 to 2000 mOsm/kg H2O |
Accuracy2 | 0 to 400 mOsm/kg H2O: mean value ≤ 2 mOsm/kg H2O from nominal value(1 SD) > 400 to < 1500 mOsm/kg H2O: ≤ 0.5% mOsm/kg H2O from normal value (1 SD) > 1500 to 2000 mOsm/kg H2O ≤ 1% mOsm/kg H2O from normal value (1 SD) |
Precision2 | (within run) 0 to 400 mOsm/kg H2O: standard deviation ≤ 2 mOsm/kg H2O > 400 to < 1500 mOsm/kg H2O: coefficient of variation ≤ 0.5% mOsm/kg H2O > 1500 to < 2000 mOsm/kg H2O: coefficient of variation ≤ 1% mOsm/kg H2O |
Temperature effects3 | Less than mOsm/kg H2O per 5 °C (9 °F) ambient temperature change |
Communication | Onboard printer, USB 2.0 Type A port(2), USB 2.0 Type B Port(1), Ethernet 10/100, RJ45 Connector port(1) |
Supported languages | Simple Chinese, Czech, Danish, English, French, German, Greek, Italian, Japanese, Korean, Portuguese, Russian, Slovak, Spanish, Swedish, Turkish |
Storage temperature | -20 °C to +45 °C (-4 °F to +113 °F) |
Electrical voltage | 100 to 240 VAC (50/60 Hz) |
Power consumption | 60 Watts |
Dimensions (D x W x H) | 38 cm x 36 cm x 38 cm (15″ x 14″ x 15″) |
Net weight | 6.0 kg (13.3 lbs) |
Shipping weight | 11.4 kg (25 lbs) |
Warranty | One-year limited warranty on workmanship and all parts except glass, Plastic, and part warranted by their makers |
- Subject to change
- Accuracy and precision (within run) specifications apply to Advanced Instruments standards
and reference solutions. Performance at Reference Conditions: 20 °C to 25 °C (68 °F to 77 °F);
40 to 60% relative humidity - Operating Conditions: Temperature 18 °C to 35 °C (64 °F to 95 °F); 5 to 80% relative humidity
(non-condensing) - Dimensions when Micro-Sample Test Kit is on the instrument