- Products
- Applications & Industries
- Bioprocessing
-
-
Compliance & Conformity
Our portfolio of instruments, consumables and services are designed to help you maintain regulatory guidelines, ensure data integrity, and meet GMP requirements.
-
- Clinical
-
-
Compliance & Conformity
Our portfolio of instruments, consumables and services are designed to support clinical lab regulations, standards and with HIPAA data security in mind.
-
- Food & Beverage
- Bioprocessing
- Service & Support
- Resources
- Company
Sample type | Aqueous-based solution |
Sample volume per test | 40 μL |
Test time | 140 seconds |
Sample Capacity | Multi plate |
Resolution | 1 mOsm/kg H2O |
Range1 | 0-2000 mOsm/kg H2O |
Accuracy2 (AI standards and reference solutions) |
Mean value ±3 mOsm/kg H2O from nominal value between 0 and 400 mOsm/kg H2O (1 SD) Mean value ±1% from nominal value greater than 400 to 2000 mOsm/ kg H2O (1 SD) |
Within-run Repeatability2 (AI Standards and Reference Solutions) |
Standard deviation ≤3 mOsm/kg H2O between 0 and 400 mOsm/kg H2O Coefficient of variation ≤1% greater than 400 to 2000 mOsm/kg H2O |
Operating conditions | Temperature: 18 °C to 35 °C (64 °F to 95 °F) Humidity: 5 to 80% relative humidity (non-condensing) |
Supported Languages | English, Simplified Chinese, French, German, Italian, Japanese, Korean, Portuguese, Russian, Spanish, Turkish |
Storage temperature | -20 °C to +45 °C (-4 °F to +113 °F) |
Electrical voltage | 100 to 240 VAC (50/60 Hz) |
Power consumption | 60 Watts |
Dimensions (D x W x H) | 48.9 cm D x 62.87 cm W x 68.58 cm H (19.25” D X 24.75” W X 27 H”) |
Net weight | 59.7 kg (131.6 lbs.) |
Shipping weight | 111.9 kg (246.6 lbs.) |
Warranty | One-year limited warranty on workmanship and parts |
- Advanced Instruments completes testing from 0 mOsm/kg sample to 2000 mOsm/kg sample during in-house linearity testing.
- Performance at reference conditions: 20 °C to 25 °C (68 °F to 77 °F); 30 to 60% relative humidity.