BR2 Bilirubin Stat-Analyzer | Advanced Instruments
We are open and continue to provide essential lab supplies during the COVID-19 pandemic. Answers to your FAQs can be found here. Read More

BR2 Bilirubin Stat Analyzer

Highlighted Features
  • Small sample size requirement
  • Test results obtained in as little as 2 minutes
  • Easy operation

BR2 Bilirubin Stat Analyzer

The Advanced® BR2 Bilirubin Stat-Analyzer™ system provides fast, accurate results for both total and direct bilirubin levels in neonates requiring only a 30µL serum or plasma sample. It determines total bilirubin by spectrophotometric analysis and direct bilirubin using a modified version of the proven Malloy-Evelyn method. Because of its compact, portable design and ease of use, it is ideally suited for pediatric physician offices, satellite labs, hospital stat labs, or pediatric research laboratories.

*The Advanced BR2 Bilirubin Stat-Analyzer is only sold in the United States.

Overview

Product Features

FAST AND ACCURATE TEST RESULTS

Provides instantaneous results for total bilirubin, direct bilirubin results achieved in 2 minutes

SMALL SAMPLE SIZE

Requires only a single 30µl serum or plasma sample for both total and direct bilirubin analysis

EASY TO USE

Directions are provided directly on instrument. Test kit provides everything required to run samples immediately

PORTABLE DESIGN

Ideal solution for a physician’s office, satellite laboratory, or mobile laboratory settings

LOW COST PER TEST

Provides economic benefit and low cost of ownership to any laboratory

PROVEN RELIABILITY

Simple instrument design backed by service and support from a world-leading manufacturer of diagnostic equipment

Technical Specifications

Advanced Model BR2 Bilirubin Stat-Analyzer Specifications*
Sample Size 30µl
Sample Type Serum or Plasma
Sample Capacity Single sample, disposable cuvettes
Test Time Total: instanteneous Direct: 2 minutes
Units mg/dl Bilirubin (also available in µmol/L) S.I. units for export only; 1mg/dl bilirubin = 17.1 µmol/L
Range O to 30 mg/dl
Linearity Within ±0.2 mg/dl from straight-line from 0 to 30 mg/dl
Resolution ±0.1 mg/dl
Repeatability Within ±0.2 mg/dl over entire range
Drift Zero: < 0.2 mg/dl after 30-minute warm up Calibration: < 0.2 mg/dl 24 hr after warm up
Hemolysis No interference from HbO2 levels up to 600 mg/dl
Turbidity Neglisible interference from turbidity associated with triglyceride (TGL) levels up to 500 mg/dl
Operating Conditions Temperature 15°C to 32°C (59°F to 90°F); 5% to 80% relative humidity (noncondensing)
Storage Temperature -40°C to +70°C (-40°F to 160°F)
Electrical Voltage 100 to 120 VAC(60Hz) or 200 to 240 VAC(50Hz)
Power Consumption 15W
Dimensions 7″H x 11″W x 13″D (18cm X 28cm X 33cm)
Net Weight 9lb (4kg)
Shipping Weight 12lb (5.5kg)
Warranty One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers

Associated Products

Supplies & Accessories

B2A101Total Bilirubin Test Kit, 100 tests
B2A102Total and Direct Bilirubin Kit, 100 tests
B2A800Cuvettes Only

Services

Take the worry out of added service costs and have total peace of mind with an Advanced Care service agreement for the Anoxomat jar system.

Educational Resources

Brochure

BR2 Bilirubin Stat Analyzer

Education

What is Neonatal Bilirubin?