Advanced® Model BR2 Bilirubin Stat-Analyzer™

Advanced® Model BR2 Bilirubin Stat-Analyzer™

Advanced Instruments Bilirubin STAT-Analyzer™, Model BR2, for Total Bilirubin and Direct Bilirubin Measurement

The Advanced Instruments Model BR2 Bilirubin STAT-Analyzer™ employs the method of Evelyn-Malloy, after Henry, to measure total bilirubin and direct bilirubin in neonatal samples.

The BR2 determines bilirubin levels through spectrophotometric analysis.

  • Works with infant plasma or serum samples
  • Requires only 30 µL of sample
  • Total bilirubin measured immediately, direct bilirubin takes just two minutes.
  • Easy to use, simple setup, calibration and operation

Uses and Applications

  • Identify hyperbilirubinema
  • Chart the normal course of post-partum physiological jaundice
  • Monitor those at high risk of kernicterus
  • Determine the onset of normal hepatic bilirubin conjugating mechanism

Advanced® Bilirubin Stat-Analyzer™ Photometer Technical Specifications

Electrical

Power Requirement: 100 to 120 VAC (60 Hz) or 200 to 240 VAC (50 Hz)
Fuses (2): 250 VAC time delay (Type T): 1/4-Amp (100-130 VAC), 1/8-Amp (200-250 VAC)
Power Consumption: 15 Watts
STAT Mode: Power left on continuously
Test Capability: Total and Direct Bilirubin, both on the same serum or plasma sample

Sample Volume: 30 μL

Sample Capacity: Single sample, disposable cuvettes

Readout: Three-digit, seven-segment LED display

Units: mg/dL or μmol/L (S.I. units for export only; 1 mg/dL bilirubin = 17.1 μmol/L)

Range: 0 to 30 mg/dL

Linearity: Within ±0.2 mg/dL from zero to 30 mg/dL (straight-line method)

Resolution: ±0.1 mg/dL

Repeatability: Within ± 0.2 mg/dL over entire range

Hemolysis: No interference from HbO2 levels up to 600 mg/dL

Turbidity: Negligible interference from turbidity associated with triglyceride (TGL) levels up to 500 mg/dL

Storage Temperature: -40 to +70°C (-40 to +160°F)

Operating Temp. (ambient): 15 to 32°C (59 to 90°F)

Drift:
Zero: Less than 0.2 mg/dL/hr after 30-minute warm-up
Calibration: Less than 0.2 mg/dL/24 hr after 30-minute warm-up

Room Humidity: 5 to 80% R.H. (non-condensing)

Test Time: Instantaneous for TOTAL Bilirubin, 2 minutes for DIRECT Bilirubin

Dimensions 
18 cm x 28 cm x 33 cm
(7" H x 11" W x 13" D)

Weight
Net:  4 kg (9 lbs.)
Shipping:  5.5 kg (12.0 lbs.)

Warranty: One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers.

Certification: