Advanced Instruments Osmometer, Model 3250, for Osmolality Measurement
The Advanced Instruments Model 3250 Micro Osmometer determines the osmolality of solutions using freezing point depression (FPD). Ideal for serum, urine or other bodily fluids in clinical applications; and media or formulations in pharmaceutical, biotech or industrial applications.
Automated, single sample, designed to process a 200-250 mL sample, and two-minute test time. The Model 3250 provides the best accuracy and precision in the market today.
Advanced Instruments also supplies ControLine Products
Method – Freezing Point Depression Osmometry
Osmometry measures the total molar concentration of dissolved solids in any solution. The freezing point of a solution is depressed in direct relation to the amount of solute in solution. Click here for a more detailed explanation.
Uses and Applications – Emergency Medicine, Internal Medicine/Endocrinology, Urology/Renal Function, Surgery/Critical Care, Obstetrics/Gynecology, Pharmacy, General Quality Control. Click here for a more detailed explanation.
Advanced® Model 3250 Osmometer Technical Specifications
Electrical
Power Requirement: 100 to 130 VAC (50/60 Hz) or 200 to 250 VAC (50/60 Hz)
Fuses (2): 250V time delay (Type T): 2-Amp (100-130 VAC); 1-Amp (200-250 VAC)
Power Consumption: 150 Watts
Memory Backup: Integral lithium cell; 10-year life (typical); (not user-replaceable)
Sample Volume: 0.20 to 0.25 mL
Sample Capacity: Single sample
Readout: 20-character vacuum fluorescent display
Units: mOsm/kg H2O
Ranges
Low: 0 to 2000 mOsm/kg H2O
High: 1400 to 4000 mOsm/kg H2O
Resolution: 1 mOsm/kg H2O
Communications: On-board printer, DTE EIA-232/V.24 (RS-232) serial port and optional barcode scanner
Performance at Reference Conditions
Linearity: Less than 0.5% from a straight line
Repeatability: 0 to 400 mOsm: ±2 mOsm/kg (1 S.D.) 400 to 4000 mOsm: ±0.5% (1 S.D.)
Drift: Less than 1 mOsm/kg H2O per month
Performance Over Operating Conditions
Temperature Effects: Less than 1 mOsm/kg H2O for every 5°C (9°F) ambient temperature change
Operating Conditions Temperature: 18°C to 35°C (64°F to 95°F)
Humidity: 5 to 80% relative humidity (non-condensing)
Storage Temperature: -40°C to +45°C (-40°F to +113°F)
Start-up Time: Immediately from stand-by; under 5 minutes from power-on
Test Time: Approximately 2 minutes per sample in low range; approximately 3 minutes in high range
Dimensions
30.5 cm x 33.0 cm x 45.7 cm
(12.0" H x 13.0" W x 18.0" D)
Height With Head Up: 40.6 cm (16.0" H)
Weight
Net: 12.7 kg (28.0 lbs.)
Shipping: 17.7 kg (39.0 lbs.)
Warranty: One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers.
Certification:

Installation Class: I
Over-Voltage
Category: II
Pollution Degree: 2
Moisture Protection: IPX0 (ordinary)